14 BM-C Beamline

 

The 14-BM-C station is as a fixed-wavelength bending magnet station especially well suited for crystallography of viruses and proteins with large unit cells, as well as collection of atomic ultra-high resolution data. Fiber diffraction is also supported.

 


 

Optics

  • Bent conical Si-mirror (Rh coated)
  • Bent Ge(111) monochromator

 


 

Beam Characteristics

  • Energy range:
    fixed energy 12.668 keV (0.979Å)
    (8-14.9 keV accessible on request)
  • Bandpass:
    ΔE/E: 3.1 x 10-4
  • Beam size, horizontal x vertical (FWHM):
    0.13 x 0.34 mm2

 


 

Station

  • Sample positioning:
    Kappa diffractometer; inclined detector geometry supported
    goniometer head with remote xyz-translation
    B&W and color sample alignment camera
    automated crystal centering
  • Sample to detector distance:
    100 to 1000 mm remotely adjustable
  • Sample temperature control:
    liquid nitrogen sample cooler
  • ADSC Quantum-315 Detector

 


 

Other Equipment/Facilities Available